P2.0 (OFS): Offset for % Saturation
Measurement

Figure 30: Offset for % saturation

NOTE

This sub group appears in % saturation measurement

mode only.

From measurement mode:

1.

Press the MODE key to select % saturation mode.

2.

Press the SETUP key to enter SETUP mode.

3. Press

▲ or ▼ keys to scroll through subgroups

until you view parameter “OFS” in the upper
display.

4. Press the ENTER key. The upper display shows

the current measurement in % saturation and the
lower shows “OFS”.

5. Press

▲ or ▼ keys to offset the % saturation

measurement.

6. Press the ENTER key to confirm selection and to

move back to subgroup “OFS”. If you want to
return to measurement mode, press CAL/MEAS.

(The values shown above are for illustration purpose. Your meter may
display different values depending on the settings & your environmental
conditions)