Touchscreen Calibration

If it seems that the FrameScope Pro is not responding correctly
to the taps on the touch screen, the touch screen may need to be
calibrated to bring the taps back into alignment with the
underlying display.

Follow the following procedure to calibrate the touch screen.

1

Ensure that you have the stylus available.

2

Select Calibrate Touch Screen and press Edit on the System
Settings Network screen. The Calibration screen will be
displayed.

3

Using the stylus, press the screen as directed by the onscreen
instructions. Since the coordinate readings are very
sensitive, you should tap on the exact point of the screen as
indicated by the drawing of the stylus.

4

The FrameScope Pro will return to the System Settings
Network screen when the calibration has completed.