10. Settings

Image 10-4

The custom display function allows the selection of a user defined display function set in Factory, which is a
linear function (Gamma = 1)

10.2.2 Display function in a closed loop medical visualization system

Display function in a closed loop medical visualization system

If, however the projector is part of a typical closed loop medical visualization system (Barco Dicom

®

Theatre ), the Display function

will be set by the control & calibration system (MediCal

®

Pro) along with a color temperature forming a pre-set value.

How to select a Display function ?

1. Refer to the system manual

OSD interaction

When a preset is selected on MediCal

®

Pro , the selected Display function (as well as the color temperature ) will be marked on the

projector’s OSD menu.

Standard
DICOM
DIN
Custom

Image 10-5
Display function : active selection

The completion of the calibration procedure will in the same way be signaled by a separate logo (as well as for the color temperature).

Standard
DICOM
DIN
Custom

Image 10-6
Display function: calibrated

70

R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004