Settings – Barco R9003110 User Manual

Page 83

Advertising
background image

10. Settings

Image 10-26

Image 10-27

Performing White input balance

1. Connect the source you want to project

2. Select a white pattern (or gray scale as alternative)

3. Press MENU to activate the Toolbar

4. Press → to select the Image item

5. Press ↓ to Pull down the Image menu

6. Use ↑ or ↓ to select Input balance

7. Press → to pull down the menu

8. Use ↓ or ↑ to select White balance (image 10-28)

9. adjust the red white level (gain) on a minimal value (image 10-29)

10.adjust the blue white level (gain) on a minimal value

Note:

this minimal value is not necessary , provided that the 2 other colors are not influencing too much the color to be adjusted,
in fact the aim is to minimize the effect of the two other colors since there is a risk of reaching too soon the transition
(bright spots) due to the contribution of these two other colors signals.

11.Adjust the Green white level (gain) until bright spots appear on the screen

12.Adjust the Blue white level (gain) until bright spots appear on the screen

13.Adjust the Red white level (gain) until bright spots appear on the screen

the projected image should know be noisy neutral grey.

Image 10-28

Image 10-29

R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004

79