6 $aa1, 5 $aa0 2.6 $aa1 – B&B Electronics Network Device I-7011 User Manual

Page 27

Advertising
background image

27

I-7011, I-7018 User Manual

Rev:B1.0

2.6 $AA1

Description : Perform Zero Calibration

Syntax : $AA1[CHK](cr)

$

delimiter character

AA

address of setting module (00 to FF)

1

command for performing zero calibration

Response :

Valid Command :

!AA[CHK](cr)

Invalid Command :

?AA[CHK](cr)

Syntax error or communication error may get no re-

sponse.

!

delimiter for valid command

?

delimiter for invalid command or the calibration is not

enabled

AA

address of response module(00 to FF)

Example :

Command : $011

Receive : !01

Perform address 01 zero calibration, return success.

Command : $021

Receive : ?02

Perform address 02 zero calibration, return the calibration is

not enabled before perform calibration command.

Related Command :

Sec.2.5 $AA0, Sec.2.20 ~AAEV

Related Topics :

Sec.1.9 Calibration

Advertising