Error Codes (RH digital)

Technical data

Device

Operating Voltage Range - Rated voltage

Vac

230 ± 10% / 115 ± 10% / 100 ± 10%

Frequency

Hz

50 / 60

Power consumption (+10%) max.

W 620

Permissible duration of operation

% 100

Permissible ambient temperature

°C

+5 to +40

Permissible relative humidity % 80

Protection type acc. DIN EN 60529

IP 21

Protection class

I

Overvoltage category

II

Contamination level

2

Operation at a terrestrial altitude

m

max. 2000

Dimensions (W x D x H)

mm

250 x 160 x 100

Weight

kg 2.8

Motor

Speed range

rpm

0; 100 - 2000

Power consumption

W 20

Setting resolution (digital/basic)

rpm 50/Scale

Speed stability (no load) at rated voltage

≥ 500 rpm

%

± 10

< 500 rpm

rpm ± 100

Stirred quantity max.(H

2

O) ltr 15

Heating plate

Dimensions (Ø)

mm 135

Material

stainless steel / enameled white

Heating

Heating power (-5%/+10%) at

rated

voltage

W 600

Adjustment and display resolution

(digital/basic)

K 5/Scale

Surface temperature

°C

ambient temperature ...320

Hysteresis of heating plate at rated voltage

K

± 20

no container, center of heating plate at 100 °C
Electronic Thermometer

DIN 12878

Adjustable safety circuit

Safety temperature limit (adjustable)

°C

100 - 360

Error Cause

Effect

Solution

code

E3

Temperature inside device is too high

Heating off - Switch off device and allow to cool down.

E4

Motor or magnetic rods blockage

Heating off - Switch off device.

Motor off - Warning! Only to be carried out by authorized service personnel:

Carry out an internal test on the device to check the plug-in connector for the motor.

E11

Break in safety circuit

Heating off - Plug in contact plug (N).

- Plug in contact thermometer/temperature sensor.

- Replace faulty connecting cable, plug, or contact thermometer.

E21

Failure in safety circuit test

- Switch off and switch on again after about 1 minute.

E22

When fault is indicated again, please contact with IKA

®

service.

E29

- Set a higher safe temperature limit.

E24

Surface temperature (Temperature of contr- Heating off - Switch off device until the surface temperature of the heating plate is lower than

ol sensor) of the heating plate is higher than the

the selected safe temperature limit.

set safe temperature limit

- Set a higher safe temperature limit.

E25

Heating and switching element monitoring Heating off - Switch off device.

- Safety temperature limit > 100

°C

See also “Functional check of inactivating the safety circuit”.

-

Warning! Only to be carried out by authorized service personnel:

Carry out an internal test on the device to check the plug-in connector for the heating element.

E26

Difference between temperature of safety

Heating off - Switch off device.

sensor and temperature of control sensor

- Warning! Only to be carried out by authorized service personnel:

Control temperature > (Safety temperature + 40 K)

Carry out an internal test on the device to check the plug-in connector for the temperature sensor.

If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
- Send the device for repair, including a short description of the fault.