Error codes (rh digital) technical data – IKA RH digital User Manual
Page 15
15
Error Codes (RH digital)
Technical data
Device
Operating Voltage Range - Rated voltage
Vac
230 ± 10% / 115 ± 10% / 100 ± 10%
Frequency
Hz
50 / 60
Power consumption (+10%) max.
W 620
Permissible duration of operation
% 100
Permissible ambient temperature
°C
+5 to +40
Permissible relative humidity % 80
Protection type acc. DIN EN 60529
IP 21
Protection class
I
Overvoltage category
II
Contamination level
2
Operation at a terrestrial altitude
m
max. 2000
Dimensions (W x D x H)
mm
250 x 160 x 100
Weight
kg 2.8
Motor
Speed range
rpm
0; 100 - 2000
Power consumption
W 20
Setting resolution (digital/basic)
rpm 50/Scale
Speed stability (no load) at rated voltage
≥ 500 rpm
%
± 10
< 500 rpm
rpm ± 100
Stirred quantity max.(H
2
O) ltr 15
Heating plate
Dimensions (Ø)
mm 135
Material
stainless steel / enameled white
Heating
Heating power (-5%/+10%) at
rated
voltage
W 600
Adjustment and display resolution
(digital/basic)
K 5/Scale
Surface temperature
°C
ambient temperature ...320
Hysteresis of heating plate at rated voltage
K
± 20
no container, center of heating plate at 100 °C
Electronic Thermometer
DIN 12878
Adjustable safety circuit
Safety temperature limit (adjustable)
°C
100 - 360
Error Cause
Effect
Solution
code
E3
Temperature inside device is too high
Heating off - Switch off device and allow to cool down.
E4
Motor or magnetic rods blockage
Heating off - Switch off device.
Motor off - Warning! Only to be carried out by authorized service personnel:
Carry out an internal test on the device to check the plug-in connector for the motor.
E11
Break in safety circuit
Heating off - Plug in contact plug (N).
- Plug in contact thermometer/temperature sensor.
- Replace faulty connecting cable, plug, or contact thermometer.
E21
Failure in safety circuit test
- Switch off and switch on again after about 1 minute.
E22
When fault is indicated again, please contact with IKA
®
service.
E29
- Set a higher safe temperature limit.
E24
Surface temperature (Temperature of contr- Heating off - Switch off device until the surface temperature of the heating plate is lower than
ol sensor) of the heating plate is higher than the
the selected safe temperature limit.
set safe temperature limit
- Set a higher safe temperature limit.
E25
Heating and switching element monitoring Heating off - Switch off device.
- Safety temperature limit > 100
°C
See also “Functional check of inactivating the safety circuit”.
-
Warning! Only to be carried out by authorized service personnel:
Carry out an internal test on the device to check the plug-in connector for the heating element.
E26
Difference between temperature of safety
Heating off - Switch off device.
sensor and temperature of control sensor
- Warning! Only to be carried out by authorized service personnel:
Control temperature > (Safety temperature + 40 K)
Carry out an internal test on the device to check the plug-in connector for the temperature sensor.
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Please contact the service department;
- Send the device for repair, including a short description of the fault.