Ground connectors, 24v output, Plc inputs – INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 100: 1 ground connectors, 2 24v output, 3 plc inputs, Caution

6-2
Equipment Connections
k
in
a
26
e Ch
ap
te
r
6
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
9)
6.1.1
Ground connectors
Pin 2, 3, 18, 19 and 21 are ground connectors.
6.1.2
24V Output
Pin 6 is a +24 V common output for supplying the PLC inputs and outputs, internally
protected with fuse F4.
6.1.3
PLC Inputs
These inputs can be used to control the Protec P3000 via a programmable logic
control (PLC).
Technical data
Pin assignment (default)
All PLC inputs can be defined by the user from a list of commands. The commands
can be selected in the
SETTINGS
/
INTERFACES
/
SELECT
PLC
INPUTS
submenu. For
details on this submenu please refer to Section 4.4.4. The currently selected
commands can be viewed on INFO page #7.
Standby
Change from
LOW
to
HIGH
to activate
STANDBY
function.
Change from
HIGH
to
LOW
to deactivate
STANDBY
function (wake up).
Caution
Permissible maximum input voltage 28 V.
24V nominal input
Low level: 0 …7 V
High level: 13 … 28 V
Pin
Default Command
7
Standby
8
Clear Error
9
Calibration Abort
13
Calibration
20
Zero
25
Not used