6.1.1

Ground connectors

Pin 2, 3, 18, 19 and 21 are ground connectors.

6.1.2

24V Output

Pin 6 is a +24 V common output for supplying the PLC inputs and outputs, internally
protected with fuse F4.

6.1.3

PLC Inputs

These inputs can be used to control the Protec P3000 via a programmable logic
control (PLC).

Technical data

Pin assignment (default)

All PLC inputs can be defined by the user from a list of commands. The commands
can be selected in the

SETTINGS

/

INTERFACES

/

SELECT

PLC

INPUTS

submenu. For

details on this submenu please refer to Section 4.4.4. The currently selected
commands can be viewed on INFO page #7.

Standby

Change from

LOW

to

HIGH

to activate

STANDBY

function.

Change from

HIGH

to

LOW

to deactivate

STANDBY

function (wake up).

Caution

Permissible maximum input voltage 28 V.

24V nominal input

Low level: 0 …7 V

High level: 13 … 28 V

Pin

Default Command

7

Standby

8

Clear Error

9

Calibration Abort

13

Calibration

20

Zero

25

Not used