Extract the air filter.
4
Depending on the contamination the air filter should be cleaned or replaced as
necessary.
5
Push in new (changed) air filter until fully in contact with dead stop.
6
Close air filter fastener.
7
From the main menu go to
HISTORY
&
MAINTENANCE
/
CONFIRM
MAINTENANCE
and
confirm with the appropriate button on the right side of the display that the air filter
has been replaced / cleaned. The actual date and time as well as the actual number
of operating hours will be saved and a reminder to perform this maintenance task
again will be issued after another 10.000 operating hours.
Fig. 7-2 Dismounting of the air filter
Fig. 7-3 Inserting the air filter