Extract the air filter.

4

Depending on the contamination the air filter should be cleaned or replaced as

necessary.

5

Push in new (changed) air filter until fully in contact with dead stop.

6

Close air filter fastener.

7

From the main menu go to

HISTORY

&

MAINTENANCE

/

CONFIRM

MAINTENANCE

and

confirm with the appropriate button on the right side of the display that the air filter
has been replaced / cleaned. The actual date and time as well as the actual number
of operating hours will be saved and a reminder to perform this maintenance task
again will be issued after another 10.000 operating hours.