Electrical Data

Other data

1.2

Support by INFICON

INFICON Service

If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health or whether
it is contaminated. If it is contaminated also indicate the nature of the hazard.
INFICON must return any equipment without a Declaration of Contamination to the
sender’s address. You will find an appropriate form below.

Gas flow through the capillary

For Protec P3000

260 - 360 sccm*

For Protec P3000XL in

HIGH

FLOW

mode

2660 - 3500 sccm*

Time until ready for operation

approx. 3 min

* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased
altitude and low atmospheric pressure.

Mains voltages and mains frequencies

90 - 127 V, 50 / 60 Hz

(fixed)

115 - 140 V, 60 Hz

187 - 265 V, 50 / 60 Hz

Power consumption

200 VA

Type of protection

IP 20

Overvoltage category

II

Mains cord

2.5 m

Noise level

< 54 dBA

Dimensions (w x h x d) in mm

610 x 370 x 265

Weight

27 kg

Permissible ambient temperature
(during operation)

10 °C to 45 °C

Permissible storage temperature

-40 °C to 60 °C

Relative humidity

max. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C

Contamination level

II
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”)

Max. altitude above sea level

2000m

Advertising