INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 98

Advertising
background image

5-8

Protec P3000 Messages

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

* To be performed only by authorized INFICON service personnel.

W86

Internal calibration not possible

Not in measure mode when
calibration is attempted

Wait until Protec P3000 enters
measurement mode!

W87

Not supported gas in PRO-
Check

Occurs only during internal
calibration or proof function:

ECO-Check installed
(accidentially)

EEPROM in PRO-Check not
programmed

Take out ECO-Check, install PRO-
Check!

Replace PRO-Check reference leak!

W88

PRO-Check defective

Occurs only during internal
calibration or proof function:
Temperature sensor defective

Replace PRO-Check reference leak!

W89

Overrange!

Protec P3000 is contaminated with
helium

Do not switch off Protec P3000(XL)!!!
Keep Protec P3000 running while
providing fresh air to sniffer line until
warning is cleared.

If warning occurs frequently, increase
contamination limit!

Check helium background

See Section

4.6

, Info page 2.

Test leak value too high during
external calibration!

Use smaller test leak for external
calibration!

W90

Calibration conditions not
maintained

Sniffer removed during internal
calibration, internal calibration
aborted

Repeat calibration acknowledge

Error
no.

Plain text message

Possible Reason

Recommendation