Equipment Settings

4-20

k

in

a

26

e Ch

ap

te

r

4

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Name

Each program can be named with a free text, descriptive name up to 6 characters.
The name can be edited by highlighting the name line and pressing

EDIT

. An

alphanumerical entry screen will open. Press ok after entering the desired name
description.

Gas type

For editing the gas equivalent go to the “

GAS

” line item and press “E

DIT

”. Only gas

equivalents that are listed in the “

GAS

/

TRIGGER

” submenu (Gas1 to Gas4) are

available for selection. The gas equivalent is stated in plain text in parenthesis. The
list of maximum four available gas equivalent can be scrolled with the up and down
buttons.

Notice

The I•Guide Mode refers to the gas equivalent selected in the “

GAS

/

TRIGGER

” submenu. If you change a gas equivalent in the “

GAS

” submenu

the gas equivalent selected for the I•Guide program will automatically
change as well.

No. of points

The number of potential leak locations to be checked can be varied between 0 and
99 and can be edited with the up and down keys.

Default value: 4

Notice

The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Protec P3000 will prompt for the next location to be
tested continuously without using the global leak rate function.

Notice

The I•Guide mode can also be used to summarize leak rates on demand.
If the number of points is set to 99, a results screen with summarized global
leak rate will be displayed after the right button has been pressed for 2 s
continiously (or after the 98th point automatically).

Measuring time

The

MEASURING

TIME

during which the sniffer tip needs to be hold to the correct

location can be programmed to values between 0.7 and 25.0 s. The measuring time
should not be less than the response time of the Protec P3000. We recommend to
use the following minimum measuring times:

Please use the

UP

and

DOWN

buttons for editing this parameter.

Default value: 1.0 s

Sniffer line length

min. measuring time

3m

0.7 s

5m

0.9 s

10m

1.4 s

15m

3.0 s