Controls on the probe display unit, Performing measurements, 3 controls on the probe display unit – INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 39: 4 performing measurements, Warning

Advertising
background image

Operation of the Protec P3000

3-5

k

in

a

26

e Ch

ap

te

r

3

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

3.3

Controls on the probe display unit

On the display of the probe handle similar information as on the main display is
shown.

The currently detected leak rate is indicated as a bar graph. In a second line the
numerical leak rate (in the same unit of measurement as on the main display) is
shown. In the third line the type of gas equivalent is stated (e.g. He or R134a).

For Protec P3000XL only:
If the Protec P3000XL is operated in

HIGH

FLOW

mode, the gas equivalent is

displayed in inverted characters (on dark background).

The sniffer probe offers two press buttons. By pressing the left button the current
background reading is set to ZERO. For details on the ZERO function please refer to

4.4.1

. ZERO.

The right probe button is used for navigation when working in I•Guide mode (see
Section

3.4.2

) or for starting an internal calibration (see Section

3.5.2

)

3.4

Performing measurements

The Protec P3000 offers two modes of operation:

The Standard Operation Mode (compatible to the Protec mode)

The I•Guide Operating Mode

Standby Function

The Protec P3000 offers a

STANDBY

function to prevent the needless intake of

contaminants into the sniffer probe during idle times, thus saving filter and sensor life
time.

Pos.

1

2

3

Description

Bar graph indicating the leak rate

Absolute leak rate

Gas equivalent

Fig. 3-5 Sniffer display in standard operation mode

1

3

2

Warning

Danger of electric shock.

Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.

Advertising