Protec P3000 Messages

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

E55

Flow through sniffer below error
limit

Only for Protec P3000 or Protec
P3000XL in

LOW

FLOW

mode:

Flow through sniffer line below limit
low flow error for

LOW

FLOW

mode

Filter in sniffer line blocked!

Internal filter in main unit blocked

Limit low flow error for

LOW

FLOW

mode too high

Replace sniffer tip filter and recalibrate!
(see section 7.4)

Replace internal filter* and recalibrate!

Check limit low flow error (for

LOW

FLOW

mode)! See section 4.4.1

E56

HIGH

FLOW

below error limit

Only for Protec P3000XL in

HIGH

FLOW

mode:

Flow through sniffer line below limit
low flow error for

HIGH

FLOW

mode

or pressure at Wise Technology
sensor lower than corresponding
lower pressure limit

Filter in sniffer line blocked!

Internal filter in main unit blocked

Limit low flow error for

HIGH

FLOW

mode too high

Replace sniffer tip filter and recalibrate!
(see section 7.4)

Replace internal filter* and recalibrate!

Check limit low flow error (for

HIGH

FLOW

mode)! See section

4.4.1

W59

EEPROM parameter queue
overflow!

May occur if software “update” to
older version is performed

Restart the Protec P3000, if problem
persists, call nearest INFICON service
representative!

W60

All EEPROM parameter lost!
Please check your settings!

New EEPROM has been installed,
EEPROM on motherboard is virgin

All settings in software menu are reset
to default! Please enter your settings
again!

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

W61

EEPROM parameter initialized! Software

update performed and

new parameters have been
introduced

Newly introduced parameters are
listed below warning

Acknowledge warning

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

W62

EEPROM parameter lost!

Parameter has been modified
during software update and reset
to default value

Affected parameters are listed
below warning

Check setting of modified parameters in
corresponding software menu and set
to desired value!

If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty

Replace EEPROM*

W64

There are outstanding
warnings!

Acknowledged but still valid
warnings will be repeated every 2
hrs or on each new power on

Please double-check the warnings!

Error
no.

Plain text message

Possible Reason

Recommendation