Flow settings
(For Protec P3000XL only)
The Protec P3000XL can be used in
LOW
FLOW
mode as well as in
HIGH
FLOW
mode.
In the flow settings submenu all settings related to both flow modes can be adjusted.
When entering the
FLOW
SETTINGS
submenu four further submenus will be displayed:
Flow limits
LOW
FLOW
In this submenu the
UPPER
FLOW
LIMIT
and
LOWER
FLOW
LIMIT
for the
LOW
FLOW
mode
may be set-up. The
LOWER
FLOW
LIMIT
can be set to values between 150 and 240
sccm. Default value: 180
The
UPPER
FLOW
LIMIT
can be varied between 300 and 650 sccm. Default value: 395
Fig. 4-10 Changing the flow limits of the SL3000 sniffer line