By pressing the right probe button the next testing cycle may be started.

Notice

The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Protec P3000 will prompt for the next location to be
tested continuously without using the global leak rate function.

Notice

The I•Guide mode can also be used to summarize leak rates on demand.
If the number of points is set to 99, a results screen with summarized global
leak rate will be displayed after the right button has been pressed for 2 s
conintously (or after the 98th point automatically).

3.4.3

The Info Page

By pressing the info button on the main display a general info page will open.
Information on the software version used, date & time information, the currently set
audio volume and the minimum audio volume are stated.

If an active warning exists this will be displayed in the info page instead of the gas
info lines.

Fig. 3-13 Result of I•Guide program: unit under test failed

Fig. 3-14 Info page without errors or warnings