7 maintenance, Maintenance schedule, 7maintenance – INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 106: 1 maintenance schedule, Warning

Advertising
background image

7-1

Maintenance

k

in

a

26

e Ch

ap

te

r

7

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

7

Maintenance

7.1

Maintenance schedule

Required tools

2 screw drivers, size 2

1 ring spanner wrench, 19 mm

8 mm hexagonal screw driver (delivered with Protec P3000).

Key for Maintenance Schedule:

I Repair level I Customer

II Repair level II Customer with technical training from INFICON

III Repair level III INFICON service engineer

1) Depends on the environment

Warning

For all maintenance on the Protec P3000, the main power must be disconnected
first.

Required maintenance

Sub

Assembly

Material

Description

Part no.

Operation hours

Repair
level

2000 5000

10.000

Check sinter filter and
replace if necessary

Sniffer tip

Sinter Filter for
Sniffertip SL3xx,
SL3000-x (5 pcs.)

200 03 500

X

I

Replace if W35 „Flow too
low“

Sniffer tip

Felt for Capillary Filter SL3xx,
SL3000-x (50pcs.)

200 001 116

1)

I

Check internal filter and
replace if necessary

Main Unit

internal Filter (10 pcs.)

200 001 680

X

II

Maintenance of
diaphragm pump

Diaphragm
Pump
MVP015

Wearing parts kit for
Diaphragm pump

200 03 504

X

III

Clean or replace the air
filter at the bottom of the
chassis

Main Unit

Air Filter Protec P3000
(104x154 mm; 5 pcs.)

200 001 552

X

I

Replace gas reservoir
after 1 year

PRO-Check

Replacement gas reservoir
for PRO-Check

521-010

1)

I

Replace filter pad

Sniffer tip

SL3000-XL

200 002 251

I

Advertising