INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 67

Advertising
background image

Equipment Settings

4-12

k

in

a

26

e Ch

ap

te

r

4

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Flow mode (only available when SL3000XL sniffer line is connected)

In this submenu the Protec P3000XL can be switched between

LOW

FLOW

mode and

HIGH

FLOW

mode.

Default setting:

HIGH

FLOW

Mode

Limits low flow error

This submenu is only available if the

ADDITIONAL

LOW

FLOW

ERROR

is set to

ENABLED

.

In this submenu the limits for the

ADDITIONAL

LOW

FLOW

ERROR

may be set-up. If the

actual flow through the sniffer line drops below the

LIMITS

LOW

FLOW

ERROR

, the

Protec P3000XL will stop measuring. Only when the cause of the low flow problem
has been resolved and the flow is above the

LIMITS

LOW

FLOW

ERROR

again, the

Protec P3000XL will go back into measurement mode.

Two different values for the HIGH FLOW and LOW FLOW mode may be entered.
The value for the HIGH FLOW mode will also be converted into a corresponding
LOWER PRESSURE LIMIT. If the corresponding LOWER PRESSURE LIMIT is
undercut, this is an indication of the smaller capillary's flow being blocked in
SL3000XL sniffer line and also E55 / E56 will be issued and the Protec P3000XL will
stop measuring.

Error limits can be set to values between 150 and 240 sccm for LOW FLOW mode
and to values between 1500 and 2400 sccm for HIGH FLOW mode.

Sniffer button

In this submenu the use of the right sniffer button for switching between

LOW

FLOW

and

HIGH

FLOW

mode during leak testing (except when in I·Guide mode) may be

ENABLED

or

DISABLED

. If the use of the

SNIFFER

BUTTON

is

DISABLED

, the Protec

P3000XL may only be switched from

HIGH

FLOW

to

LOW

FLOW

mode (or vice versa)

through the

FLOW

MODE

submenu or via RS232.

Default setting: enabled

Fig. 4-13 Additional limits for low flow error

Default value:

150 sccm for LOW FLOW mode

1500 sccm for HIGH FLOW mode