Equipment Settings

4-26

k

in

a

26

e Ch

ap

te

r

4

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

Printer manual

The

PRINTER

MANUAL

protocol allows to send leak rates on pressing the right sniffer

probe button.

Notice

When in any of the two printer modes, no ASCII commands must be sent
to the Protec P3000 as this will interrupt the printer mode.

When set to I·Guide mode the PRINTER MANUAL mode is without function.

Printer auto

In

PRINTER

AUTO

the leak rate will be sent any time the set trigger level is exceeded.

At the time the signal drops below the trigger level the maximum detected leak rate
will be sent.

When set to I·Guide mode the leak rate results will be sent after each test point
together with the test point number. After the test cycle has been completed, the
summarized global leak rate will be sent.

Format

Date

Time

Gas equivalent Leak rate

Unit

Example

03.05.2005

12:31

He

6x10

-4

mbar l/s

Format

Date

Time

Gas equivalent Leak rate

Unit

Example

03.05.2005

12:31

R134a

2.34

g/a

Example:

Point no.

Date

Time

Leak rate

Unit

Point 01

8.3 E-6

mbar l/s

Point 02

2.0 E-6

mbar l/s

Point 03

8.4 E-6

mbar l/s

Point 04

1.2 E-6

mbar l/s

Global

13.04.2007

11:57:03

1.8 E-6

mbar l/s