INFICON Protec P3000(XL) Helium Leak Detector User Manual
Page 81

Equipment Settings
4-26
k
in
a
26
e Ch
ap
te
r
4
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
9)
Printer manual
The
PRINTER
MANUAL
protocol allows to send leak rates on pressing the right sniffer
probe button.
Notice
When in any of the two printer modes, no ASCII commands must be sent
to the Protec P3000 as this will interrupt the printer mode.
When set to I·Guide mode the PRINTER MANUAL mode is without function.
Printer auto
In
PRINTER
AUTO
the leak rate will be sent any time the set trigger level is exceeded.
At the time the signal drops below the trigger level the maximum detected leak rate
will be sent.
When set to I·Guide mode the leak rate results will be sent after each test point
together with the test point number. After the test cycle has been completed, the
summarized global leak rate will be sent.
Format
Date
Time
Gas equivalent Leak rate
Unit
Example
03.05.2005
12:31
He
6x10
-4
mbar l/s
Format
Date
Time
Gas equivalent Leak rate
Unit
Example
03.05.2005
12:31
R134a
2.34
g/a
Example:
Point no.
Date
Time
Leak rate
Unit
Point 01
8.3 E-6
mbar l/s
Point 02
2.0 E-6
mbar l/s
Point 03
8.4 E-6
mbar l/s
Point 04
1.2 E-6
mbar l/s
Global
13.04.2007
11:57:03
1.8 E-6
mbar l/s