Flow display (only available
in
ADVANCED
mode)
In the
FLOW
DISPLAY
submenu the continuous display of the flow through the sniffer
line during measurements can be
ENABLED
or
DISABLED
. If
ENABLED
, the flow will be
indicated in the
STANDARD
OPERATION
MODE
only underneath the leak rate bar graph
as integer number.
Notice
After certain events like acknowledging warnings or when switching the
flow mode, the flow monitoring may be disabled for a short period of time
while the flow stabilizes to a final value again and no new warnings will be
issued. During this time a message ”Flow control inactive” will be displayed
instead of the flow value.
Default setting: disabled
User mode
In this submenu either the
NORMAL
or the
ADVANCED
mode can be selected. In the
NORMAL
mode only the most relevant information for normal operation will be
displayed whereas in the
ADVANCED
mode additional information for advanced users
will become available.
Fig. 4-16 Setting-up the peak hold function