Audio Functions

In the Audio sub-menu all settings regarding the different audio alarms of the Protec
P3000 can be edited.

Beep sound

Beep sounds are sounds that are issued as confirmation of certain functions, e.g.
completion of a calibration process. This acknowledgement sounds can be enabled
or disabled in the beep sound submenu.

Default value: enabled.

Audio intern

The internal loudspeaker of the main unit can be switched on and off. This will not
effect the headphones outlet.

Default value: enabled.

Audio sniffer

The loudspeaker of the probe handle can be set to issue an alarm when the

TRIGGER

level is exceeded, when the

SEARCH

level is exceeded or can be disabled completely.

Default value:

TRIGGER

.

Alarm type

Different types of audio alarm of the main unit speaker can be selected. Options are

SETPOINT

,

PINPOINT

, and

TRIGGER

ALARM

.

Default value: Trigger Alarm

SETPOINT

The frequency of the tone will change when the trigger level is
exceeded. Below the trigger level the loudspeaker will remain off.

TRIGGER

ALARM

An acoustic alarm with a continuous low tone will sound as soon as
the search level is exceeded. An acoustic alarm which consists of
two tones sounds as soon as the trigger level is exceeded. Below the
search level the loudspeaker will remain off.

If the Protec P3000 is set to TRIGGER ALARM as the alarm type an
additional button will be shown on the lower right side called TONE
1,2,3. With this button different alarm sounds can be selected for the
trigger alarm to clearly distinguish between nearby other leak
detectors.

PINPOINT

The external loudspeaker is always on. Within the leak rate range of
+/- one decade of the trigger level the frequency will change
depending on the leak rate. Outside this range the tone will be
constantly low or high. This setting is recommended if also leaks
which are still below the trigger level need to be pinpointed or if the
expected leak rates are within the range of the trigger level.