INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 94

Advertising
background image

5-4

Protec P3000 Messages

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

E37

Flow through sniffer too high

Actual flow is higher than upper
flow limits (

LOW

FLOW

and

HIGH

FLOW

mode) or actual pressure is

higher than upper pressure limit
(

HIGH

FLOW

mode only):

Capillary broken or leaky

Upper flow limit setting too low

Replace sniffer line and recalibrate
or
Replace probe cable*

Increase upper flow limit
See Section

4.4.1

E38

Pump error

Fore pump defective

Call nearest INFICON service
repesentative!

W39

Valve block error!

Cannot recognize valve block
version during self-test

Call nearest INFICON service
representative!

W40

Ratio of

HIGH

FLOW

to

LOW

FLOW

failed

For Protec P3000XL only in

HIGH

FLOW

mode:

Flow in

HIGH

FLOW

mode less than

5 times flow in

LOW

FLOW

mode

(during start-up or calibration only)

Filter in sniffer line blocked

Internal filter in main unit blocked

Leak in

LOW

FLOW

capillary

Internal leak

Replace sniffer tip filter and
recalibrate(See Section

7.4

)

Replace internal filter* and recalibrate

Replace sniffer line and recalibrate
If problem persists, call nearest
INFICON service representative

W41

HIGH

FLOW

too low

For Protec P3000XL only in

HIGH

FLOW

mode:

Actual flow lower than lower flow
limit (for high flow mode) or actual
pressure at Wise Technology
sensor below corresponding
pressure limit

Filter in sniffer line blocked

Internal filter in main unit blocked

Leak in

LOW

FLOW

capillary

Internal Leak

Replace sniffer tip filter (See Section

7.4

)

Replace internal filter*

Replace sniffer line!
Or: Replace probe cable!*

See Section

7

Error
no.

Plain text message

Possible Reason

Recommendation