INFICON Protec P3000(XL) Helium Leak Detector User Manual

Page 97

Advertising
background image

Protec P3000 Messages

5-7

k

in

a

26

e Ch

ap

te

r

5

.fm

Tech

ni

ca

l Ha

nd

bo

ok

(1

20

9)

W65

Wrong date set!

Wrong date in the Protec.

Wrong Hex code entered to
PRO-Check

Check the date in the Protec.

Check the Hex code entered to
PRO-Check

W66

New PRO-Check!

New PRO-Check reference leak
installed in main unit

Please enter serial number and code.

See Section

4.7

, Replace PRO-Check

W67

PRO-Check will be expired on
DD.MM.YYYY

Warning for pending expiration will
be issued adjusting 14, 30, 60, or
90 days before actual expiration.

Please order new PRO-Check gas
reservoir! (cat. no. 521-010)

See Section 7.7.5

W68

PRO-Check expired!

PRO-Check has been used for
more than 1 year or has been
manufactured more than 2 years
ago.

Please replace PRO-Check gas
reservoir!

See Section

7.7

W70

All EEPROM parameter of
PRO-Check lost!

EEPROM in PRO-Check is empty
or faulty

Replace PRO-Check!

W71

No communication with PRO-
Check!

No or defective electrical
connection between PRO-Check
and main unit

Check connection of PRO-Check with
main unit
If problem persists, call nearest
INFICON service repesentative!

PRO-Check not installed in main
unit

Install PRO-Check!

No PRO-Check available

Disable PRO-Check in software menu
(see Section

4.5.6

)

W72

No communication with sniffer!

No or defective electrical
connection between sniffer line
and main unit

Check connection of sniffer line with
main unit (disconnect and reconnect, if
possible, try different sniffer line)
If problem persists, call nearest
INFICON service

W78

Signal difference between
calibrated leak and air too
small!

Calibrated leak too small / empty
during calibration or proof

Check leak rate of test leak or use leak
with higher leak rate

Background too high during
calibration

Check helium background

See Section

4.6

, Info page 2.

Not enough time expired for air
signal to stabilized (acknowledged
too early)

Repeat calibration, allow sufficient time
for air signal to stabilize!

W81

Calibration factor too low!

Calibration factor determined to be
< 0.1 during calibration

Leak rate of test leak is incorrect
(especially during external
calibration)

Check correct setting of test leak value!

W82

Calibration factor too high!

Calibration factor during calibration
determined to be > 10

Leak rate of test leak is incorrect
(especially during external
calibration)

Check correct setting of test leak value!

Test leak not sniffed properly / not
long enough

Repeat calibration, sniff test leak
properly and for sufficient time

Error
no.

Plain text message

Possible Reason

Recommendation