The Info Menu
The
INFO
MENU
lists all internal data that may be useful for any trouble-shooting of the
Protec P3000. The
INFO
MENU
consists of 10 pages. On each page the upper left
button “
BACK
” allows to leave the info menu, the lower left button allows to go back
one page whereas the lower right button allows to go to the next page. The page
number is displayed in the upper right corner.
Menu item
Format
Description
Page 1: General Data
Foreline pressure
mbar
Flow
sccm
Flow through sniffer line
Time since power on
Min
Operation hours
h
Serial number
11-digit number
Wise Serial number
Software version
x.xx.xx
Circuit temperature
°C
Temperature of main board
Test leak temperature
°C
Sensor block
Protec P3000 1 / 2 / 3
1 = Protec P3000
2 = Protec P3000 with
extended measurement
range
3 = Protec P3000XL
Page 2: Sensor Global
I akt.
A
B-Filter
A
I-Filter
A
Q-Filter
A
Leak rate
mbar l/s
helium leak rate
PWM-Membrane
Integer
Heater Voltage
V
High Voltage
V
Wise-State