Main Power Switch (I/O) and Emergency-Stop

MTS Criterion

®

Series 40 Product Manual

Operation

103

Main Power Switch (I/O) and Emergency-Stop

Main power switch

(I/O)

The main power switch is located on the right side of the frame base. Select
position I to turn the power on to the load frame and the controller. When the
power is on, the displays on the frame control panel illuminate. Select position O
to turn off power to the load frame and the controller.

3-Phase Mains Power

Switch

C45.305/C45.605 load frame also has a 3-Phase mains power switch which is
located on the right-back side of the frame base. The 3-Phrase main power switch
needs to be turned on before the main power switch (I/O) will work properly.

Emergency Stop

The frame is also equipped with an Emergency Stop button. The
Emergency-Stop will cut the power to the motor and should be used for
emergency purposes only.

The Emergency-Stop should be periodically pressed when the controller is
powered but not when a test is running. The controller continously monitors the
redundant Emergency-Stop chain and will generate a fault alerting the user if
any problems are detected. Pressing Emergency-Stop allows the active state to
be checked.

Criterion_product manual.book Page 103 Monday, July 28, 2014 3:52 PM