Controller Connections

MTS Criterion

®

Series 40 Product Manual

Installation

69

Controller Connections

Connecting the Main Power

1 kN through 100 kN

The input voltage of MTS Criterion frames rated 100 kN or less is single phase
200-230 V, 50/60 Hz.

Note

Local electrical codes supercede any information found here.

For MTS Criterion frames equal or less than 10 kN, use 3 holes type I socket
(C13 style on the Criterion) for power input, specification of power wire is

H05VV-F,3G1 mm

2

, compatible with standards of KEMA-KEUR, CEBEC,

+S+S+S, VDE, SABS, IEMMEQU for certification.

For MTS Criterion frames more than 10 kN, use the 3-wires cable that is
provided for power input, and connect to the customer electrical box and
disconnect. The type of power cable is listed below:

10 kN~50 kN - HO7RN-F 3G2.5 mm

2

, compatible with standards of ELOT,

PECSO for certification.

100 kN - HO7RN-F 3G4 mm

2

, compatible with IEMMEQU, PECSO for

certification.

Note

Electrical connections must be made by qualified personnel and is their
responsibility for using the proper power disconnect that conforms to
local codes and regulations when connecting the machine to the
building’s main power.

300 kN through 600kN

The input voltage of MTS Criterion frames rated 300kN or more is three phases
380-480 V, 50/60 Hz.

For MTS Criterion frames more than 10 kN, use the 3-wires cable that is
provided for power input, and connect to the customer electrical box and

disconnect. The type of power cable is : HO7RN-F 3G4 mm

2

, compatible with

IEMMEQU, PECSO for certification.

Electrical disconnect

The customer is responsible for providing an electrical power disconnect that is
easy to operate and easy to reach. It must also meet IEC 60947-1 and IEC 60947-
3 standards.

Recommended circuit breaks would be ones that are of the thermal magnetic type
with characteristics suitable for large inductive loads (D-type trip characteristic).
If fuses are used, It is recommended that they are of the time-delay type with dual
elements. These recommendations should be followed to avoid nuisance tripping.