Installing Cables

MTS Criterion

®

Series 40 Product Manual

Installation

75

J11 Guard

The connector should be connected to the safety system (Enclosure switch) or
jumpered for crosshead motion.
The enclosure switch has two NC contacts. One must be connected between Pin
1 and Pin 6. The other must be connected between Pin 2 and Pin 7.
GUARD_CONFIG1/2-jumpers determine if crosshead motion slows to 540 mm/
min or comes to a complete stop when the enclosure door opens. That is to say,
No Jumpers = complete stop with door open.
Jumpers from Pin3-Pin8 and Pin5-Pin9 = crosshead slows to 540 mm/min with
the enclosure door open.

Pin assignments are as follows:

P

IN

S

IGNAL

1

GUARD1A+

2

GUARD2A+

3

GUARD_CONFIG1+

4

No Contact

5

GUARD_CONFIG2-

6

GUARD1A -

7

GUARD2A-

8

GUARD_CONFIG1-

9

GUARD_CONFIG2+