Common specifications, Common specifications 36 – MTS Criterion 40 - Manual User Manual

Page 36

Advertising
background image

MTS Criterion

®

Series 40 Product Manual

36

Specifications

Introduction

Common Specifications

The following specifications are for all MTS Criterion frames. Specifications for
the specific models are located in the following tables.

C

ALIBRATION

S

TANDARD

ISO 7500 C

LASS

0.5

OR

ASTM E4

ISO 7500 C

LASS

1

Force range

1-100% of rated force capacity

0.5-1% of force rated
capacity

Rated force capacity at max test speed

100%

Maximum test speed at rated force
capacity

100%

Force indicating accuracy

± 0.5 % of indicating

Speed accuracy

Set speed

< 0.01 mm/min: speed accuracy is within ± 1.0% of set

speed

Set speed

≥ 0.01 mm/min: speed accuracy is within ± 0.2% of set

speed

Position accuracy

Within ± 0.5%

Strain accuracy

ASTM E83 or ISO 8513

Security protection

Over-Force, travel limits, over-voltage and others

Over force protection

10%

Data acquisition rate

1000

Hz

Control loop rate

1000

Hz

Environmental requirements

Operating temperature

Operating humidity

Storage temperature

Maximum storage humidity

Maximum altitude

For indoor use only

5-40°C (41-104°F)

5-85% Noncondensing

-18-49°C (0-120°F)

90% Noncondensing

2000 m (6562 ft)

Motor and

drive system

AC Servomotor

Ball

screws

Preloaded

Position measurement

Encoder

Additional DC conditioning channels

2 channels

Additional incremental encoder
conditioning channels

1 channel

Criterion_product manual.book Page 36 Monday, July 28, 2014 3:52 PM