J8 and j9 dc conditioner, J10 aux pw1 – MTS Criterion 40 - Manual User Manual
Page 74
![background image](/manuals/568017/74/background.png)
MTS Criterion
®
Series 40 Product Manual
74
Installing Cables
Installation
J8 and J9 DC
Conditioner
Two external DC conditioner connectors are provided. Up to two external
transducers can be connected such as: axial or transverse extensometers, biaxial
extensometer, auxilary load cells, pressure gages, LVDTs (with external
conditioning), or strain-gaged components.
Pin assignments are as follows:
J10 Aux Pw1
The connector is the auxilary Power connector. It can provide power to external
devices. Pin assignments are as follows:
P
IN
S
IGNAL
1
TEDS data
2
EX+
3
EX-
4
FB-
5
RCAL1 (FBR+)
6
RCAL2 (FBR-)
7
FB+
8
EXS-
9
EXS+
10
TEDS ground
P
IN
S
IGNAL
1
+12 V
400 mA (max combined with Pin 6)
2
AGND
3
-12 V
400 mA max
4
PGND
5
+5 V
400 mA max
6
+12 V
400 mA (max combined with Pin 1)
7
AGND
8
DGND
9
+24 V
400 mA max
Criterion_product manual.book Page 74 Monday, July 28, 2014 3:52 PM