Installing Cables
Installation
J14 E- Stop
This connector is intended to connect to a remote Emergency-Stop switch. If not
used, a jumper plug must be installed. If you are building a cable, the maximum
length is 30.48 m (100 ft) with 18 AWG. The switch should be wired normally
closed, such that when the switch is pressed, an interlock is generated. Pin
assignments are as follows:
•
ESTOP3 = emergency stop switch with two sets of contacts (optional
remotely mounted).
•
ESTOP_OUT = optional ESTOP output for accessory, includes two contact
outputs and one feedback. For use with pneumatic/hydraulic grip controls,
furnaces, environmental chambers, and so on.
P
IN
S
IGNAL
1
ESTOPB_OUT-
2
No Contact
3
ESTOP_OUT_MONITOR-
4
ESTOP_OUT_MONITOR+
5
ESTOP3B+
6
ESTOPA_OUT-
7
ESTOP3B-
8
ESTOP3A+
9
ESTOPB_OUT+
10
No Contact
11
No Contact
12
No Contact
13
ESTOP3A-
14
ESTOPA_OUT+
15
No Contact