Installing Cables

Installation

J14 E- Stop

This connector is intended to connect to a remote Emergency-Stop switch. If not
used, a jumper plug must be installed. If you are building a cable, the maximum
length is 30.48 m (100 ft) with 18 AWG. The switch should be wired normally
closed, such that when the switch is pressed, an interlock is generated. Pin
assignments are as follows:

ESTOP3 = emergency stop switch with two sets of contacts (optional
remotely mounted).

ESTOP_OUT = optional ESTOP output for accessory, includes two contact
outputs and one feedback. For use with pneumatic/hydraulic grip controls,
furnaces, environmental chambers, and so on.

P

IN

S

IGNAL

1

ESTOPB_OUT-

2

No Contact

3

ESTOP_OUT_MONITOR-

4

ESTOP_OUT_MONITOR+

5

ESTOP3B+

6

ESTOPA_OUT-

7

ESTOP3B-

8

ESTOP3A+

9

ESTOPB_OUT+

10

No Contact

11

No Contact

12

No Contact

13

ESTOP3A-

14

ESTOPA_OUT+

15

No Contact