286.31 Transient Permeability Pore Pressure Intensifier

16

Purging and Filling the Fluid Lines

Operation

Purging and Filling the Fluid Lines

In this section you’ll purge the 286.31’s lines of air and fill them with distilled
water.

You only need to do this after installing the 286.31 or if you’ve accidentally
drawn air into the Pressure Intensifier through a dry Fluid Reservoir.

Note

Before you start purging and filling, make sure your Triaxial Cell is open.

Purging and filling has ten major tasks, which must be done in order

Fill the Fluid Reservoir

Purge and Fill the Pressure Intensifier

Pressurize the Intensifier

Purge and Fill the Lower Pore Line

Purge and Fill the Reference 1 Lines

Purge and Fill the Balance and Upper Pore Lines

Purge and Fill the Reference 2 Lines

Purge and Fill the Low Pressure Reservoir Line

Purge and Fill the Reservoir Return Line

Refill the Pressure Intensifier