Running a Steady State Flow Test

286.31 Transient Permeability Pore Pressure Intensifier

Operation

49

Purge and Fill the Pore Pressure Lines

1. Purge and fill the lower pore pressure line.

Turn the Pore Fluid Input valve to the Open position.

As soon as bubble-free fluid comes out at the triaxial cell’s lower pore