Purge and fill the pore pressure lines, Purge and fill the pore pressure lines 63 – MTS Model 286-31 Transient Permeability Pore Pressure Intensifier User Manual

Page 63

Advertising
background image

Running a Transient Permeability Flow Test

286.31 Transient Permeability Pore Pressure Intensifier

Operation

63

Purge and Fill the Pore Pressure Lines

1. Purge and fill the lower pore pressure line.

Turn the Pore Fluid Input valve to the Open position.

As soon as bubble-free fluid comes out at the triaxial cell’s lower pore
pressure port, go on to step 2.

Advertising