Prepare to run the test, Prepare to run the test 70 – MTS Model 286-31 Transient Permeability Pore Pressure Intensifier User Manual

Page 70

Advertising
background image

286.31 Transient Permeability Pore Pressure Intensifier

70

Running a Transient Permeability Flow Test

Operation

\

Prepare to Run the Test

1. Install your specimen in the triaxial cell.

For the details on this procedure, refer to the installation section of your
triaxial cell manual.

If pore pressure becomes higher than confining pressure — even for a
second — your specimen will be damaged.

Always keep confining pressure higher than pore pressure.

2. Apply a confining pressure to the specimen.

Always keep confining pressure higher than pore pressure.

For details on this step, refer to the operation section of confining pressure
intensifier's manual.

3. If your triaxial cell has pore pressure valves, make sure they are open.

CAUTION

Advertising