Description

Item

Moves enabled channels to a predefined Ride level prior to test startup.

Ride

7.3.2.0 Manually Position the Actuator to Install and Remove
Specimens

This procedure is used for installing and removing specimens.

1. If desired, set up the Scope window to view the actuator’s displacement and force signals.

2. On the Station Manager window’s Station Controls panel toolbar, click the Manual Command button.

3. Set up the Manual Controls window:

a) Select the channel to be manually commanded.
b) Select the desired control mode.
c) Select the Enable Manual Command check box.

Warning:

Applying hydraulics can result in sudden actuator motion.

A moving actuator can injure anyone in its path.

Always clear the actuator area before applying hydraulics.

4. In the Station Manager window’s Station Controls panel:

a) Click Reset to clear any interlocks.
b) As needed, click Reset/Override to override interlocks.

Each Reset/Override click adds 10 seconds of override time.

c) Apply station hydraulic power.

When possible, use low pressure. As soon as power is applied, the Manual Command window’s Manual
Cmd shows the actual actuator position.

5. In the Manual Command window, ensure that the Enable Manual Command check box is selected.

6. If you are using a handset, select the Exclusive Control check box.

7. Use the Manual Cmd slider bar to position the actuator, or select the value shown, enter a new value, and

click Enter on the keyboard.

Multipurpose Elite Test Design Guide 101