0 running the test procedure – MTS Multipurpose Elite Test User Manual

Page 54

Advertising
background image

Value

Property

Signal List

Note:

Click the Signal List plus sign and use the
Signals Selection window to select the
following signal options:

Ch 1 Displacement Ch 1 Force Ch 1 Output

Buffer Size

Note:

Click the Advanced button to set the
buffer size.

100 count

5.2.12.0 Running the Test Procedure

In this task, run the completed test procedure.

1. Ensure that the station has power by setting the HPU to high. If an interlock indicator is lit on the Power

panel, click Reset. Next, click the Low and then High buttons on the HSM.

2. Click the New Test Run icon in the tool bar.

3. In the Specimen Selection window, note the generic specimen selection and click OK.

4. Observe the Setup Variables window, and then click OK.

5. Observe the MTS TestSuite Input Variables window. For End Level 1, enter 3 mm. For End Level 2,

enter -1 mm. When you are finished, click OK.

6. To start the test, click the Run button (green arrow) on the Program control panel.

7. Observe the test run on the Signal Scope. The signal trace ramps to a mean level of 4 mm, performs five

cycles with a true sine waveshape to end levels of 7 mm and 3 mm, and returns to zero.

54 Multipurpose Elite Test Design Guide

5.0 Creating Tests