0 channel data syntax – MTS Multipurpose Elite Test User Manual

Page 89

Advertising
background image

Suggestions

Comments

Keyword

This is an optional but recommended entry.

Placed in the file by the Profile
Editor to determine the
dimension of a given unit.

Dimension=

6.2.11.7.0 Channel Data Syntax

Following the channel definition, you must define the command data for that channel.

Channel data attributes can be entered two ways, as constants or as individual values. Constants are declared
before any individual parameters are listed, and they require the equal sign (=) to be appended to the end of
its keyword. Any keyword can be defined as a constant. The following is one example of the command data
syntax where one keyword is defined as a constant. Each row of data defines a command element.

The following is an example of proper command data syntax:

Command Data Syntax

Shape= Sine

Level2

Level1

Count

Frequency

mm

mm

Cycles

Hz

-5

5

100

10

0

10

150

5

0

0

0.5

1

Channel Data Syntax

Comments

Range

Keyword

This entry is required. Only one of these three
can be specified. Rate must reflect the rate of
change of the level dimension.

Undetermined

Frequency, Time, or

Rate

1

This optional entry specifies the number of
repeats. The units can be segments or cycles,

Greater than 0

Count

and the value must be greater than 0. If this
option is omitted, the command will run once.

This entry is optional. When not assigned as
a constant, each segment can have a different
shape. The default is Sine.

Ramp, Sine, Square, TrueSine,
SquareTapered, RampTapered,
SineTapered, or
TrueSineTapered

Shape

1

Rate is not supported by phase profiles.

Multipurpose Elite Test Design Guide 89

6.0 Creating Advanced Tests

Advertising