101 • Instrument Function Menu

Option 2 Sub Tab

Bump Test Function: The Bump Test Function setting is used to turn the bump
test function on or off. If deselected (factory setting), the BUMP menu item will
not appear in the Calibration Mode menu.

NOTE:

The following parameters can be changed and uploaded to the
instrument even if Bump Test Function is deselected and these
settings are not displayed or used at the instrument.

Bump Test Time Remaining: With Bump Test Time Remaining selected, the
instrument will give an indication at start up if it is due for bump testing. The type
of indication depends on the Bump Test Past Due Act setting. With Bump Test
Time Remaining deselected (factory setting), the instrument will give no
indication at start up if it is due for bump testing.

Bump Test Interval(Day): The Bump Test Interval(Day) setting defines the
amount of time between bump tests. The minimum setting is 0 days and the
maximum setting is 30 days (factory setting).

Bump Test Past Due Act: The Bump Test Past Due Act determines how the
instrument will react to a needed bump test during the start up sequence. If

Figure 77: Parameter Tab, Option 2