RKI Instruments SDM-2009 PC Controlled Configuration User Manual

Page 42

Advertising
background image

37 • Connecting Instruments to the PC Controller Program

3.

Install a GX-2009 in the instrument cradle.

NOTE:

If the GX-2009 is left in the SDM-2009 after bump testing or calibration,
the pump will run for 12 seconds every 2 minutes in order to keep
hydrogen generated from charging of the NiMH batteries from affecting
the CO sensor.

4.

Turn on each SDM-2009 by pressing and holding the SDM-2009’s POWER
button. When the control panel LEDs turn on, release the POWER button.

5.

The BUMP

T

and CAL

S

LEDs will turn off and the COPY LED will be

steadily on or off. It will be on if the SDM-2009 was used in Standalone
Configuration and calibration and bump test records are still stored in the
SDM-2009’s memory (see the SDM-2009 Docking Station Standalone
Configuration Manual for more information). The CHARGE LED will be
blinking green if the SDM-2009 is operating properly or solid red if there is a
system failure.

6.

When a connection has been established between the SDM-GX Docking
Station PC Controller Program and all of the SDM-2009s, the main program
window will appear as shown below.

Figure 26: SDM-2009 Connection

7.

Press and hold the POWER MODE button on the instrument until you hear a
beep, then release it. The instrument will begin its power up sequence (see

Advertising