65 • Charging an Instrument in an SDM-2009

Charging an Instrument in an SDM-2009

The SDM-2009 can be used to charge the rechargeable batteries in a GX-2009. To
maximize the instrument’s run time and the battery life, make sure the batteries’
charge is as low as possible before recharging them. You can use the GX-2009’s
REFRESH function to discharge the batteries. See the GX-2009 Operator’s Manual
for a complete description of the REFRESH button.

Recharging the Batteries After Performing a Bump Test
or Calibration

To recharge the batteries after performing a bump test or calibration:

1.

Perform a bump test or calibration on an instrument as described in “Bump

• No connection

between
instrument and PC
Controller
Program

• SDM-2009 is not turned

on.

• The instrument is not

correctly inserted into the
instrument cradle.

• The IR window is dirty.

• The USB cable is not

properly plugged into the
SDM-2009 or the PC.

1. Turn on the SDM-2009. If it does

not turn on, check that the AC
adapter is plugged into an AC
socket and to the jack on the
back of the SDM-2009.

2. Check to make sure the

instrument is inserted properly.

3. Clean the IR window on the

SDM-2009 and the GX-2009.

4. Check to make sure the USB

cable is plugged into the back of
the SDM-2009 and to a USB port
on your PC.

5. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• The batteries in

the instrument are
not charging

• The instrument is not

properly inserted in the
instrument cradle.

• The charger is

malfunctioning.

• The batteries are faulty.

1. Be sure that the instrument is

fully inserted in the instrument
cradle.

2. Try to charge the batteries in a

BC-2009 charger if one is
available to determine if the
SDM-2009 is at fault.

3. Try to charge the batteries using

a different SDM-2009 if one is
available to determine if the
SDM-2009 is at fault.

4. Replace the batteries and

attempt to charge them. See the
GX-2009 Operator’s Manual for
instructions to change the
batteries.

5. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.