RKI Instruments SDM-2009 PC Controlled Configuration User Manual

Page 62

Advertising
background image

57 • Calibrating a GX-2009

7.

You will then be asked to confirm the auto calibration values and the
instruments that the calibration cylinder will be used for. Be sure the auto
calibration values match the values listed on the calibration cylinder. The serial
number and gas combination of all selected instruments will appear
sequentially beneath the gas cylinder concentration. All channels will be
calibrated using GAS 1.

Figure 42: Gas Order Confirmation Screen

8.

Click OK. You will be asked if you want to calibrate the instrument(s). Click OK
to proceed. If you don’t want to proceed with the calibration, click Cancel to
return to the main program window.

Advertising