RKI Instruments SDM-2009 PC Controlled Configuration User Manual

Page 68

Advertising
background image

63 • Calibrating a GX-2009

If at any point during the calibration the flow rate to the instrument becomes
too low, the PC Controller will abort the calibration and display a failed
calibration screen. The CAL

S

LED on the SDM-2009 control panel will be

solid red. Click OK to return to the main program window or double click the
instrument information line to view the calibration results in the Logs
window. The CAL

S

LED on the SDM-2009 control panel will turn off.

Confirm all tubing connections are correct and that all lines are clear.

Figure 48: Low Flow Calibration Failure

13. For both calibration result outcomes, the instrument(s) will be displayed in the

main program window in Details View. To return to Icons View, click the Icons
radio button.

14. When a calibration is performed, any bump test or calibration data that is being

stored in the instrument will be downloaded to the PC Controller Program. This
data can be viewed in the Logs window along with the results of the calibration
just performed. See “Bump Test Data” on page 89 or “Calibration Data” on
page 76 for instructions for viewing bump test or calibration data.

The calibration results will also be stored in the GX-2009.

Advertising