53 • Bump Testing a GX-2009

If at any point during the bump test the flow rate to the instrument becomes
too low, the PC Controller Program will abort the bump test and display a
failed bump test screen. The BUMP

T

LED on the SDM-2009 control

panel will be solid red. Click OK to return to the main program window or
double click an instrument information line to view the bump test or
calibration results in the Logs window. The BUMP

T

LED on the SDM-

2009 control panel will turn off. Confirm all tubing connections are correct
and that all lines are clear.

Figure 39: Low Flow Bump Test Failure

14. For all bump test result outcomes, the instruments will be displayed in the main

program window in Details View. To return the main program window to Icons
View, click the Icons radio button.

15. All bump test results can be viewed in the Logs window. See “Bump Test Data”

on page 89 for instructions for viewing bump test data.

If the bump test function in the GX-2009 is turned off or if the GX-2009 does
not have the bump test capability, the bump test results will only be stored in
the Logs window of the PC Controller Program. If the bump test function is
turned on the in GX-2009, the bump test results will also be stored in the GX-
2009.