8 - calibrating the module, Chapter objective, Tools and equipment – Rockwell Automation 1771-TCM,D17716.5.108 TEMPERATURE CONTROL MOD User Manual

Page 74: When to calibrate, Indicator operation during calibration, Preparing to calibrate, Calibrating the module

Advertising
background image

Chapter

8

8-1

Calibrating the Module

In this chapter, for calibrating your module, we describe:

tools and equipment
when to calibrate
calibration

methods

indicator operation during calibration
preparing to calibrate
write calibration block
read calibration block
calibration

procedure

In order to calibrate your temperature control module you will need the
following tools and equipment:

Tool or Equipment

Description

Precision Voltage Source

0-10V, 1

µ

V resolution

Precision Multimeter

10V, 1

µ

V resolution

Industrial Terminal and

Interconnect Cable

Programming terminal for PLC processors

The temperature control module is shipped factory calibrated. We
recommend that the module be recalibrated after the first 6 months. After
the first recalibration, we recommend recalibration after each 12-month
interval.

During calibration, the RUN/FLT indicator will turn to green; the
CAL/COM indicator will turn to flashing red. The indicators will remain
with these indications throughout the calibration procedure.

If need to recalibrate the module, you must calibrate the module in an I/O
chassis. The module must communicate with the PLC processor and an
industrial terminal (or compatible personal computer). Before calibrating
the module, you must enter ladder logic into the processor memory, so that
it can send (block-transfer write) calibration data to the module, and access
(block-transfer read) calibration data from the module.

Chapter Objective

Tools and Equipment

When to Calibrate

Indicator Operation During

Calibration

Preparing to Calibrate

Advertising