Metrohm 792 Basic IC User Manual

Page 102

Advertising
background image

4.4 Methods

792 Basic IC

93

Time

Start time for integration event in min.

Event

List box with integration events to be selected

(see below).

Integration events

Disable detection

Set this mode to stop detection of new peaks. If

a peak is available at the moment of the event it

is either terminated (down slope peaks) or re-

jected (upslope peaks).

Enable detection

Clear

Disable detection

mode.

Set peak start

Force the beginning of a new peak. If a peak is

available at the moment of the event it is either

rejected (upslope) or terminated (down slope).

Set peak end

Force peak end at the time of the event. Up-

slope peaks are rejected (except those born by

Set peak start

event), down slope peaks are ter-

minated.

Enable valley-to-valley

Disable valley-to-valley

Enable valley-to-valley

Disable perpendicular drop peak separation. All

peaks are considered to be baseline-separa-

ted. The bottom of the valley becomes the

baseline point.

Disable valley-to-valley

Enable perpendicular drop peak separation

(default setting).