Metrohm 792 Basic IC User Manual
Page 91
4 Operation
792 Basic IC
82
To reset the
Run
number press [Ctrl] [F8].
The
Analysis number
window is opened where
the run number for the next run can be
modified to the desired value.
Batch
X/Y: total number of started sample queues /
current analysis number in the sample queue
(read-only).
Sample
Tab
Sample
of the passport with sample information.
Info 1
First sample description (max. 256 characters).
Info 2
Second sample description (max. 256 charac-
ters).
Volume
Injected volume in PL.
Dilution
Dilution of the sample.
Vial number
Autosampler vial position to take sample from.
Amount
Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
= C
c
•
Amount
s
/
Amount
c
Internal standard amount
Concentration of the internal standard compo-
nent for relative concentration calculations.
Date/time when...
Date and time of sample collection (the default
values are equal to the date and time when the
chromatogram starts).