Metrohm 792 Basic IC User Manual

Page 113

Advertising
background image

4 Operation

792 Basic IC

104

Base

Base for calculations (

Area

or

Height

) in quantifi-

cation and calibration procedures. Indicates,

which parameter is to be used as a peak re-

sponse.

Reference channel

Channel which is used to measure area or

height (read-only).

Formula

Formula for calibration function. There are six

possible calibration dependencies for linear

and nonlinear calibration curves.

Linear through zero

function Y = k

1

X

Linear

function Y = k

1

X + k

0

Quadratic through zero

function Y = k

1

X + k

2

X

2

Quadratic

function Y = k

1

X + k

2

X

2

+ k

0

Cubic through zero

function Y = k

1

X + k

2

X

2

+ k

3

X

3

Cubic

function Y = k

1

X + k

2

X

2

+ k

3

X

3

+ k

0

Statistical weight

Weighting parameter used to weight calibration

points deviations on calculating coefficients of

the calibration formula. Apart from the uniform

weighting

1

the values

1/x

,

1/x

2

,

1/y

and

1/y

2

are

also possible.

Standard component

Name of the standard component in the case

of

Internal standard calibration

or

Tabulated calibra-

tion

method (details see on-line help).

Concentration of standard

Concentration of the standard component in

the case of

Internal standard calibration

or

Tabu-

lated calibration

method (details see on-line

help).

Update calibration

792 BASIC IC / Method / Calibration / Update

Selection of this menu item opens the

Update

window.

Retention time

Replace the expected retention times of the

components with the retention times of the cor-

responding peaks obtained in the current run.