5 Notes – Maintenance – Faults

792 Basic IC

148

x To install a new piston seal 89

89

89

89 the special tool 94

94

94

94 is used.

x First the new seal is inserted firmly in the recess of tool 94

94

94

94 by

hand (see Fig. 17B). The seal spring must be located on the

outside.

x The tool 94

94

94

94 together with the seal is then inserted in the pump

head 34

34

34

34 and the seal pressed into the pump head recess

with the aid of tool 93

93

93

93 (see Fig. 17C).

The seal surface in the pump head

34

34

34

34 must not be damaged (avoid

contact with tool)!

5 Cleaning/replacement of inlet valve

90

90

90

90

and outlet

valve

91

91

91

91

x Contaminated or blocked valves are cleaned by rinsing with

dist. water, RBS solution or acetone. The rinsing effect can be

reinforced by brief treatment in an ultrasonic bath (max. 20 s;

if longer the sapphire sphere of the valve can be damaged).

x If this does not have the desired effect, the valves can be

disassembled as shown in Fig. 18. The valve components are

pushed out with the aid of a syringe needle inserted through

the upper opening in the valve housing 95

95

95

95. The individual

components are rinsed with dist. water and/or acetone, and

the sapphire sphere cleaned with a paper towel. The valve is

then reassembled in accordance with Fig. 18. The compo-

nents of the inlet and outlet valves are identical, they are dis-

tinguished only by the positioning of the sapphire sleeve 98

98

98

98

and the ceramic holder 100

100

100

100 (see Fig. 18).

x Valves that fail to function faultlessly after such cleaning must

be replaced.

x In the reinstallation of the inlet valve 90

90

90

90 or the outlet valve 91

91

91

91

on no account must the two outwardly identical valves be in-

terchanged. To determine which valve is which, note that the

liquid flows through the pump head from the bottom up. The

flow direction of the valves can be checked simply by blowing

through the clean valve. Both valves are installed with the

black face in the direction of the pump head (see Fig. 18).

If by mistake an inlet valve

90

90

90

90 is installed instead of the outlet valve 91

91

91

91,

an extreme pressure buildup occurs within the working cylinder, which

is not detected by the pressure transducer and will destroy the piston

seal

89

89

89

89!

Advertising