2.6 High-pressure pump

792 Basic IC

27

4 Deaerate pump

x Make sure that the 6.1834.010 aspirating tubing for the high-

pressure pump has been immersed into the eluent.

x Click the

<On>

button for

IC pump

to switch on the high-

pressure pump.

x Use the syringe connected to the purge valve 31 to aspirate

air until eluent flows into the syringe.

x Click the

<Off>

button for

IC pump

to switch off the high-

pressure pump.

x Close the rotary knob on purge valve 31

31

31

31 by turning it in a

clockwise direction (see Fig. 3).

x Remove the syringe from coupling 24

24

24

24.

5 Rinse pulsation dampener (if present)

x Place a beaker beneath the column connection capillary 20

20

20

20.

x Click the

<On>

button for

IC pump

to switch on the high-

pressure pump and rinse the pulsation dampener 52

52

52

52 filled

with isopropanol for ca. 10 min with eluent.

x Click the

<Off>

button for

IC pump

to switch off the high-

pressure pump.

6 Reduce flow rate

x Reset the original flow rate under

Flow

(e.g.

0.2 mL/min

).

x Click

<Send to unit>

to send this value to the 792 Basic IC.