Metrohm 792 Basic IC User Manual

Page 107

Advertising
background image

4 Operation

792 Basic IC

98

Additionally, the

Components table

contains the following buttons:

<Add>

Add a new component (an empty column) to

the component table.

<Del>

Delete the current component from the compo-

nent table.

<Identification>

Open

the

Peak identification

window (see Identi-

fication).

<Concentrations>

Open

the

Concentration table

(see Concentration

table).

<Graphs>

Open the component window with display of

calibration curve and calibration parameters.

<OK>

Accept all changes and close the

Component ta-

ble

.

<Cancel>

Reject all changes and close the

Component ta-

ble

.

The standard method

Default.mtw

supplied with the «792 Basic IC» pro-

gram has an empty component table. All other methods contain a

component table with pre-defined components with entries for the

names and retention times.

If the output of all peaks is required in the result report then a so-

called universal component can be defined in the component table

for this purpose; this receives the value

0

in the

Time

field and no peak

is allocated to it. The name entered under

Name

(e.g.

unknown

) is then

used for all peaks for which no component can be allocated.

Peak identification

792 BASIC IC / Method / Calibration / Identification

Selecting this menu item or clicking on

<Identification>

in the component

table opens the

Peak identification

window with parameters for tuning the

peak identification procedure.

Number of components

Number of components in the component table

(read-only).

Scheme

Quick choice of identification parameters

scheme.

Standard

Set default meanings of identification parame-

ters:

Height

for reference components,

Time

for

all other components (ordinary components).

Nonstandard

Set custom meanings of identification parame-

ters.