5 Notes – Maintenance – Faults

792 Basic IC

154

3 Clean suppressor rotor

x Clean the sealing surface of new suppressor rotor 104

104

104

104

(6.2832.000) using a lint-free cloth and ethanol.

4 Insert suppressor rotor

x Insert new suppressor rotor 104

104

104

104 in suppressor holder 105

105

105

105 in

such a way that the tubing connections at the rear of the rotor

fit in the corresponding openings inside the rotor and that one

of the three holes in the rotor can be seen from below in one

of the openings of the holder.

x If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

5 Clean connection piece

x Clean the sealing surface of new connection piece 103

103

103

103

(6.2832.010) with the aid of a lint-free cloth and ethanol.

6 Insert connection piece

x Insert new connection piece 103

103

103

103 in suppressor holder 105

105

105

105 in

such a way that connection "1" is at the top and that the three

lugs on the connection piece fit in the corresponding open-

ings of the holder.

x Screw nut 102

102

102

102 onto the thread of suppressor holder 105

105

105

105

manually (do not use tools).

7 Connect and condition the suppressor

x Reconnect the suppressor to the IC system.
x Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.